PROGRAMMING
91
SPEED
Command/Query
The speed command sets the measurement speed of the instrument.
The faster the measurement speed the lower the accuracy. This
command also queries the current measurement speed.
The faster the measurement speed the lower the accuracy. This
command also queries the current measurement speed.
Syntax MAIN:SPEE:<string><^END^M>or<^J^M>
Parameter
<string> Speed
SLOW Slow
MEDI Medium
FAST Fast
Example MAIN:SPEE:SLOW<^END^M>
Set the measurement speed to slow.
Query Syntax
MAIN:SPEE?<^END^M>or<^J^M>
Return String
<string> Speed
MAIN:SPEE:SLOW<^END>
MAIN:SPEE:SLOW<^END>
Slow
MAIN:SPEE:MEDI<^END>
Medium
MAIN:SPEE:FAST<^END>
Fast
Query Example
MAIN:SPEE?<^END^M>
MAIN:SPEE:MEDI<^END>
MAIN:SPEE:MEDI<^END>
Medium measuring speed is returned.
DISPLAY
Command/Query
The display command sets the displayed measurement as a value or
as an offset from a nominal value (Delta or Delta%)
as an offset from a nominal value (Delta or Delta%)
Syntax MAIN:DISP:<string><^END^M>or<^J^M>
Parameter
LCR-800
User
Manual
92
<string> Display
VALU Unit
Value
DELP Delta
%
DELT Delta
Example MAIN:DISP:VALU<^END^M>
Set the display to Value
Query Syntax
MAIN:DISP?<^END^M>or<^J^M>
Return String
<string> Display
MAIN:DISP:VALU<^END> Value
MAIN:DISP:DELP<^END> Delta
%
MAIN:DISP:DELT<^END> Delta
Query Example
MAIN:DISP?<^END^M>
MAIN:DISP:VALU<^END>
MAIN:DISP:VALU<^END>
Currently the display is set at value.
MODE
Command/Query
The mode command sets the measurement mode of the LCR-800.
Syntax MAIN:MODE:<string><^END^M>or<^J^M>
Parameter
<string> Primary
Measurement Secondary
Measurement
RQ Resistance Quality
factor
CD Capacitance Dissipation
factor
CR Capacitance Resistance
LQ Inductance Quality
LQ Inductance Quality
factor
LR* Inductance Resistance
ZQ* Impedance Angle
*For the LCR-821 only
ZQ* Impedance Angle
*For the LCR-821 only