   Note: 
Source current flows from the 
  to COM terminals 
Figure 3-3 
Resistance Measurements 
3.5    Measuring Frequency and Period 
Frequency measurement range: 5 Hz to more than 1 MHz. 
Period measurement range: 0.2 s to less than 1 
μs. 
Input signal range: 200 mV AC to 750V AC. 
 
The instrument uses the volts input terminals to measure frequency. The AC voltage range can be 
changed with the RANGE 
  and 
  keys. However, the signal voltage must be greater than 
10% of the full-scale range. 
3.5.1    Trigger Level and Measurement Errors 
The multimeter  uses a technique  which maintains a constant resolution for any input frequency  to 
measure frequency and period. The gate time is always a multiple of the measured signal period rather 
than a fixed time.  The  error  will be no more than +/-1 from the total  gate  counts,  which  assures an 
equivalent accuracy over the whole frequency range. 
 
3.5.2    Gate Time 
Gate time is the amount of time the meter  uses to sample frequency or period readings. The 
measurement speed rate and the measuring frequency change the gate time. 
3.5.3    Connections 
Assuming  the multimeter  is under factory default conditions, the basic procedure for measuring 
frequency and period is as follows: 
1.  Connect test leads to 
  and COM terminals. 
2.  Select frequency or period measurement functions by pressing    
    or 
  →         
3.  Connect test leads to the source as shown in Figure 3-4: 
 
CAUTION:  Do not exceed 1000  V peak between 
  and  COM, or  instrument  may be 
FREQ 
FREQ