Bk Precision BK-2831E Digital-Multimeter, DMM, VFD 20000 counts CAT II 300 V BK-2831E Hoja De Datos

Los códigos de productos
BK-2831E
Descargar
Página de 78
51 
 
<b> =    1 or ON   
 
Enable auto range 
 
 
 
0 or OFF 
 
Disable auto range 
Query: 
:AUTO?   
 
 
 
Query auto range (ON or OFF) 
 
Description: 
These commands  are used to control auto ranging. With auto ranging enabled, the instrument 
automatically goes to the most ideal range to perform the measurement. 
 
The auto range command (:RANGe:AUTO) is coupled to the command that manually selects the 
measurement range (:RANGe <n>). When auto range is enabled, the parameter value for :RANGe 
<n>  changes to the automatically selected range value. Thus, when auto range is disabled, the 
instrument remains at the automatically selected range. When a valid :RANGe <n> command is 
sent, auto ranging disables. 
 
:REFerence <n> Commands 
:REFerence <n> 
Command syntax: 
:VOLTage:AC:REFerence <n>   
 
 
Specify reference for ACV 
:VOLTage:DC:REFerence <n>   
 
 
Specify reference for DCV 
 
Command parameter: 
<n> =    -757.5 to 757.5 
 
 
Reference for ACV 
-1010 to 1010            Reference for DCV 
DEFault               
0 (All measurement functions) 
          
MINimum                Minimum value for specified function 
MAXimum               Maximum value for specified function 
Query: 
:REFerence?                  
Query reference for relative function 
 
Description: 
These commands are used to establish a reference value for the specified function. When reference 
 
is enabled (:REFerence:STATe), the result will be the algebraic difference between the input signal 
 
and the reference value: 
 
Reading = Input signal – Reference 
 
From the front panel, reference is called relative (REL). 
 
The  :REFerence <n> command is coupled to the :ACQuire command. The last command sent 
(:REFerence <n> or :ACQuire) establishes the reference. When a reference is set using 
the :REFerence <n> command, the REFerence? query command returns the programmed value. 
Conversely, when a reference is set using the  :ACQuire  command,  the  :REFerence?  query 
command returns the acquired reference value. 
(Note:  In order to send a REFerence? query, the reference value must be set either