h) transistor test “hfE”
 
 transistor tests can only be performed with the optionally available measuring adapter. No volt-
age may be applied and measured on the adapter.
-  Turn the DMM on and select measuring range “hFE”. 
-  Disconnect all measuring leads from the measuring instrument.
-   Attach the optionally available measuring adapter to the three measuring 
sockets COM (5) + V (8) + mA (7).
-   Insert the transistor to be tested into the corresponding header observing 
the correct polarity. The left header is intended for NPN types and the right 
header for PNP types. SMD types can also be tested.
-  The amplification factor “hFE” is displayed.
-   After measuring, remove the adapter and turn the DMM off. Turn the rotary 
switch to the position “OFF” or turn the device off via the “POWER” switch.
i) temperature measuring (only vC150-1)
 
 the enclosed temperature sensor is designed for a temperature range from -40 to +230 °C, which 
is sufficient for most applications. to be able to use the full measuring range of the device, you 
need an optional k-type sensor. here you might need to use the optional measuring adapter.
-  Turn the DMM on and select measuring range “°C”.
-  Disconnect all measuring leads from the measuring instrument.
-   Connect the thermal sensor provided and/or the optional measurement adapter to the DMM. Make sure that the 
connection is correct (i.e. the poles are the right way around). Turn the plug so that the sensor port marked “COM” 
(-) is connected to the “COM” (5) socket and the sensor port marked “°C“(+) is attached to the “°C” (7) socket.
-  Now expose the sensor tip to the temperatures. 
-   The display shows the temperature on the sensor. If “I” appears, the measuring range was exceeded or there is 
no sensor connected.
-   After measuring, remove the adapter and turn the DMM off. Turn the rotary switch to the position “OFF” or turn the 
device off via the “POWER” switch.
   If the two sockets “COM” (5) and “°C” (7) are short-circuited, the ambient temperature of the measuring 
device is displayed.
 
 The use of K-type sensors with miniature plugs requires the application of the optional measuring adapter 