Operating and maintenance advice 
  The battery has to be taken out, when the device is stored above 50°C. 
  Handle the device and the sensor with care and only use it according its 
specification. The connectors have to be saved from humidity and dirt.  
  Unplug the sensor not by pulling the cable, but the plug.  
Tip: It is recommended to take the battery out, when the device is stored for a 
longer period of time. 
 
Operation 
Display- and Operation elements 
 
1.  display of the oxygen concentration in % or of 
MOD in meters(n) or feet(F)  
2.  BAT-warning: battery is used up and has to be 
replaced (measuring is valid) 
 
 
On/Off switch  
 
Press shortly: change between  
                    : actual value % oxygen 
: hold value % oxygen,                      
: hold value MOD in meters(n)  
                                                                  or feet(F) 
 
Press shortly: short display of the sensor rating (calculated after the last 
calibration) (i.e. “100.P“ = 100%) 
Press long:     start calibration (see also chapter 8 Calibration) 
First start  
If not already done: plug the sensor to the device. 
Screw flow-diverter on sensor, plug on transparent T-piece 
Switch the device on with the
 
 key. 
Start the calibration by pressing the
 
 key for 2 seconds, until 
 is 
displayed. After the calibration the device is ready for measuring.  
(see also chapter 8 Calibration) 
Display when switching-on the device 
 
Start and display test 
 
If switch-
off delay “P.oF” (power off) is active, this is signalled at the turn-on 
procedure.