Greisinger GOX 100 T Oxygen Meter with Sensor 600222 Hoja De Datos

Los códigos de productos
600222
Descargar
Página de 8
H59.0.22.6C-03 
 
page 5 of 8 
Oxygen measuring 
Switch on instument, calibrate (refer to chapter 8 Calibration) 
-> the current oxygen concentration [%] will be displayed 
hold sensor to valve like shown (good 
contact necessary, otherwise there 
could be measured to low values) 
Open valve a little, until You can hear 
gas flow.  
Attention: Do not open to far! 
When measurement display is stable, 
press 
 -key: 
 
hld (Hold = measurement is held) and 
oxygen concentration [%] are displayed alternating 
pressing 
 again: 
 
The Maximum Operating Depth (MOD) is displayed alternating to hld 
(n is the symbol for display in meters, F for feet) 
pressing 
 again: return to current oxygen concentration display 
 
The oxygen measurement depends on air pressure. The current pressure can be entered 
during configuration.  
If you measure at atmosphere and the calibration took place at the same conditions, the 
pressure value has not to be adjusted necessarily, because the same pressure is used for 
calibration and measuring.  
(If the measurement is carried out at other pressure conditions than the calibration: The 
actual pressure has to be entered in the configuration for  measurement and calibration. ) 
The 
sensor‟s temperature dependency is compensated internally. However for best results 
try to keep the same temperature both during calibration and measuring. 
Temperature differences between sensor and gas may cause additional errors! 
Therefore wait an appropriate period of time until the sensor has adjusted its temperature to 
the gas that is measured. A suitable flow of the gas around the sensor increases the 
adjustment significantly.  
Try to avoid warming of the sensor by touching it during measurement or calibration. 
When measuring bottled gas, consider that the gas coming out is cooler than the ambient 
because of the pressure loss! Strong air flows can produce an over pressure at the sensor 
– 
possible source of measuring/calibration errors! 
The nominal life time can be shortened by: 
Wrong storage / operation temperature 
Permanent use with dry gases (compressed gas). 
It helps to put the sensor to normal-
humid ambient air in measuring breaks (“flush” 
system with fresh air). 
The optimum operation position is: with the sensor inlet pointing downwards