miniLAB 1008 User's Guide  
Functional 
Details 
Caution! 
Port A0 through Port C7 have no overvoltage/short circuit protection. Do not 
exceed the voltage limits or you may damage the pin or the miniLAB 1008. To 
protect these pins, you should use a series resistor.  
19
37
1
20
 
Figure 4-2. DB37 Digital I/O Connector 
Table 4-2. DB37 Connector Pin-Out  
Pin 
Signal Name 
Pin 
Signal Name 
n/c 
20 
USB +5 V 
2 n/c 
21 
GND 
Port B7 
22 
Port C7 
Port B6 
23 
Port C6 
Port B5 
24 
Port C5 
Port B4 
25 
Port C4 
Port B3 
26 
Port C3 
Port B2 
27 
Port C2 
Port B1 
28 
Port C1 
10 
Port B0 
29 
Port C0 
11 GND 
30 Port 
A7 
12 n/c 
31 Port 
A6 
13 GND 
32 Port 
A5 
14 n/c 
33 Port 
A4 
15 GND 
34 Port 
A3 
16 n/c 
35 Port 
A2 
17 GND 
36 Port 
A1 
18 n/c 
37 Port 
A0 
19 
GND 
  
  
Refer to the "
" section for descriptions of cables that are 
compatible with the DB37 digital I/O connector. 
Screw terminal wiring 
The miniLAB 1008 has two rows of screw terminals. Each row has 15 connections. Pin 
numbers are identified in Figure 4-3.  The pins are labeled for eight-channel single-
ended mode operations. 
4-4