5 Memory / 6 Miscellaneous
27
GB
5.2 Deleting memory
Press M-button       to enter the memory recall mode.                                                 
 i
If you only want to delete the last measured value, press and hold the SET / SNOOZE - 
button       for approx. 7 seconds. The last measured value will flash two times and will 
then be deleted. The display shows "dEL". Press START/STOP            , to exit the de-
lete mode. If you want to delete all stored values in memory for a user, press and hold 
M-button       AND 
       simultaneously for approx. 10 seconds. 
SET / SNOOZE - button
The display shows "dEL ALL". Now press the 
       to confirm 
SET / SNOOZE - button
the deletion. The display shows "dEL OK" - now all stored values for this user are de-
leted and the unit will switch off automatically. If you want to abort the deletion process, 
press START/STOP          while the display shows "dEL ALL". The unit then switches 
itself off without deleting any value.
6.1 Error messages and error remedying
Error indicators:
6 Miscellaneous
Symbol      Cause
Correction
+ Lo
Low battery 
Bluetooth transfer of
data not possible
Cuff not properly
fitted
Cuff too tight
Failure during
inflation
Movement or speaking 
during measurement or 
pulse too weak
Pulse cannot be
detected
Measurement failed
Calibration error
The built-in battery is too low or empty.
Charge the battery.
Check the software and if Bluetooth is properly 
configured and activated on the receiving units.
Wrap the cuff properly. Remeasure in correct 
way.
Wrap the cuff properly. Remeasure in correct 
way.
Wrap the cuff properly. Check, if the air hose
is properly connected to the unit. Remeasure
in correct way.
remeasure in correct 
Rest 30 minutes and 
way. Do not speak or move during the
measurement.
Wrap the cuff properly. Remeasure in correct 
way.
Rest 30 minutes and remeasure in correct 
way.
Wrap the cuff properly. Remeasure in correct 
way. 
+ Err
E1
E2
E3
E10/E11
E20
E21
EExx