Bacharach PCA 3 Manual De Propietario

Descargar
Página de 100
PCA
3 Manual 
 
 
0024-9472 Rev 2 
 
43 
90% Efficiency 
Condensing Furnace 
80% Efficiency Fan 
Assist or Power Vented 
Furnace 
Atmospheric/Gravity 
Vented Boiler 
 
Figure 5-4.  90% 
Efficiency Condensing 
Furnace
 
 
Figure 5-5.  80% 
Efficiency Fan Assist or 
Power Vented Furnace
 
 
Figure 5-6.  
Atmospheric/Gravity 
Vented Boiler
 
Condensing 
furnaces/boilers can be 
tested through a hole 
drilled in the plastic vent 
pipe (when allowed by the 
manufacturer or local 
authority of jurisdiction) or 
taken from the exhaust 
termination. 
Combustion testing of fan 
assist or power vented, 
furnaces/boilers should be 
done through a hole drilled 
in the vent immediately 
above the inducer fan. 
Boilers, which have a ‘bell’ 
shaped draft diverter on 
top, should be tested 
directly below the diverter 
through a hole drilled in the 
vent connector. 
 
 
5.5.  Performing a Combustion Test 
 
Ensure that the following items have been completed. 
• 
Turn ON analyzer and allow it to warm up (see page 41) 
• 
Select fuel being burned (see page 23) 
• 
Inset probe into stack (see page 42) 
• 
If necessary, insert the optional primary air thermocouple into the 
combustion air stream of the burners that use an outside source of 
combustion air. 
Next, proceed with the combustion test as described below.