NOTE:  The calibration range is from –6 to –2 inwc (–15 to 
-5 mb).  An attempt to calibrate outside this range will 
cause the message “Applied Value High” (or Low) to 
appear at the bottom of the screen. 
 
  
 
Figure 6-2.   Pressure Sensor Calibration
 
7.  Wait until the measured reading stabilizes, and then press ENT to 
calibrate the pressure sensor’s Measured value to that of the applied 
value; after which the message “Good Calibration” should briefly appear 
followed by the CALIBRATION LIST screen being re-displayed. 
8.  Remove calibration equipment. 
 
Figure 6-3.   Pressure Sensor Calibration Equipment
 
 
6.5.  T-Stack Calibration 
 
This procedure first zeros and then spans the stack temperature channel to 
known temperature values.