The use of an electronic thermocouple simulator is the preferred method of 
producing the desired calibration temperatures. Alternatively, containers of ice 
water and boiling water can be used. 
6.5.1.  Material Required 
• 
Thermocouple Simulator (K-type): 
Range:  
0 to 600 °F 
Accuracy:   ±0.5 °F 
• 
(Alternatively) Ice Water, Boiling Water, Thermometer 
 
6.5.2.  TS-Zero Procedure 
1.  Set thermocouple simulator to room temperature and plug its output into 
the T-STACK connector located at the bottom of the analyzer.  
Alternatively:
 Plug the probe’s thermocouple into the T-
STACK connector located at the bottom of the analyzer. 
 
CAUTION:  DO NOT attach the probe’s gas hose to the 
analyzer’s GAS port or water will be drawn into the 
analyzer! 
2.  If not already done, turn ON the analyzer and display the 
CALIBRATION LIST screen (refer to Calibration on page 62). 
3.  Use the 
▲▼ buttons to highlight T-Stack, and then press ENT to 
display the CALIBRATE TS-ZERO screen. 
 
NOTE:  “Measured” is the current temperature reading, 
while “Applied” is a known temperature that will be applied 
for calibration purposes. 
4.  Set thermocouple simulator to 32 °F (0 °C), and then use the 
▲▼ and 
◄► buttons to enter an applied value that exactly equals the setting of 
the simulator. 
Alternatively: 
Submerge probe tip into an ice-water bath 
with a thermometer, wait several minutes, and then use the 
▲▼ and ◄► buttons to enter an applied value that exactly 
equals the thermometer reading.
 
 
NOTE:  The calibration range is from 32 to 41 °F (0 to 5 
°C).  Calibrating outside this range will cause the message 
“Applied Value High” (or Low) to appear at the bottom of 
the screen. 
5.  Wait until the measured reading stabilizes, and then press ENT to 
calibrate the TS-Zero Measured value to that of the applied value; after