Bacharach PCA 3 Manual De Propietario

Descargar
Página de 100
PCA
3 Manual 
 
 
70 
 
0024-9472 Rev 2 
6.7.2.  Procedure 
1.  If not already done, turn ON the analyzer and display the 
CALIBRATION LIST screen (refer to Calibration on page 62). 
2.  Use the 
▲▼ buttons to highlight CO-LO, and then press ENT to display 
the CALIBRATE CO screen. 
 
NOTE:  “Measured” is the current CO reading, while 
“Applied” is a known CO level that will be applied for 
calibration purposes. 
3.  Attach a 500
 ppm CO cylinder to the regulator of the calibration fixture 
4.  Use the 
▲▼ and ◄► buttons to enter an applied value that exactly 
equals the concentration stamped on the CO cylinder. 
 
NOTE:  The calibration range is from 9 to 1,500
ppm. An 
attempt to calibrate outside this range will cause the 
message “Applied Value High” (or Low) to appear at the 
bottom of the screen. 
 
   
 
Figure 6-7.   CO
LOW
 Sensor Calibration
 
5.  Adjust the regulator for a flow meter indication of approximately 2
 SCFH. 
Wait until the measured reading stabilizes (approximately 3
 minutes), 
and then press ENT to calibrate the CO Measured value to that of the 
applied value. The message “Good Calibration” should briefly appear 
followed by the TEST GAS CO/H
2
 screen. 
 
NOTE:  If the sensor’s output is low, but still usable, then 
the message “Good Calibration WARNING Low Sensor”  
will appear. The sensor will now be marked as being Low 
in the DIAGNOSTICS screen.