NOTE:  A CO-only calibration can be done by pressing 
F3
 in Step 5 
(instead of 
ENT
). 
 
6.8.  SO
2
 Sensor Calibration 
 
This procedure spans the optional sulfur dioxide sensor to a known gas
level. 
6.8.1.  Material Required 
• 
Calibration Kit 
( P/N 0024-7059) 
• 
Gas Cylinder 
(100 ppm SO
2
 in Nitrogen, P/N 0024-1158) 
 
6.8.2.  Procedure 
1.  If not already done, turn ON the analyzer and display the 
CALIBRATION LIST screen (refer to Calibration on page 62). 
2.  Use the 
▲▼ buttons to highlight SO
2
, and then press ENT to display 
the CALIBRATE SO
2
 screen. 
 
NOTE:  “Measured” is the current SO
2
  reading, while 
“Applied” is a known SO
2
  level that will be applied for 
calibration purposes. 
3.  Attach a 100
 ppm SO
2
 cylinder to the regulator of the calibration fixture 
4.  Use the 
▲▼ and ◄► buttons to enter an applied value that exactly 
equals the concentration stamped on the SO
2
 cylinder. 
 
NOTE:  The calibration range is from 9 to 1,800
ppm. An 
attempt to calibrate outside this range will cause the 
message “Bad Calibration Wrong CAL Entry” to appear in 
the following step. 
 
   
 
Figure 6-9.   SO
2
 Sensor Calibration
 
5.  Adjust regulator for a flow meter indication of approximately 2
 SCFH. 
Wait until the measured reading stabilizes (approximately 3
 minutes),