PCA
3 Manual 
 
 
78 
 
0024-9472 Rev 2 
CHAPTER 7.  MAINTENANCE 
 
 
7.1.  Overview 
 
Customer maintenance of the PCA
®
3 is limited to the following: 
• 
Battery replacement (see page 20) 
• 
Sensor re-calibration (see page 62) 
• 
Water trap / filter assembly maintenance (see pages 49 and 83) 
• 
Sensor replacement (see page 84) 
• 
Cleaning the probe (see page 87). 
All other maintenance should be performed by an authorized Bacharach Service 
Center. Refer to Service Centers on page 95. 
 
7.2.  PCA
®
3 Disassembly 
 
The following procedure describes and illustrates how to disassemble the 
analyzer. 
 
7.2.1.  Tools Required 
 
• 
Medium Phillips Screwdriver 
7.2.2.  Procedure 
1.  Unplug all thermocouples from bottom of analyzer. 
2.  Remove battery cover and then remove batteries. 
 
NOTE:  In Step 3, if the sensors are not being replaced, leave the 
tubing connected to each sensor’s gas cap, being careful not to put 
unnecessary strain on the tubing during the disassembly process. 
3.  Pull off sensor retainer; and then unplug all sensors. 
4.  Lay analyzer face down on a padded work surface; and then using a 
medium Phillips screwdriver, remove the unit’s four rear-case screws. 
5.  Lift rear case from analyzer and set aside. 
6.  Unplug electrical connectors J8, J9, and J14 from printed circuit board. 
 
 
CAUTION:  In Step 7, note that there is tubing connected between 
the bottom hose connector plate and the pressure sensor on the 
printed circuit
board. Do not put unnecessary strain on this tubing 
during the disassembly process. 
7.  Carefully lift battery and pump chassis, along with the bottom hose 
connector plate, from analyzer.  
8.  Lift printed circuit board from analyzer.