Teledyne Tablet Accessory T803 Manual De Usuario
Communications Setup and Operation
Teledyne API T803 CO2/O2 Analyzer Operation Manual
112
6.3.2. MANUALLY CONFIGURING THE NETWORK IP ADDRESSES
There are several circumstances when you may need to configure the interface
settings of the analyzer’s Ethernet card manually:
settings of the analyzer’s Ethernet card manually:
your LAN is not running a DHCP software package
the DHCP software is unable to initialize the analyzer’s interface
you wish to create a static IP (recommended)
the DHCP software is unable to initialize the analyzer’s interface
you wish to create a static IP (recommended)
1.
Connect a cable from the analyzer’s Ethernet port to a Local Area Network
(LAN) or Internet port.
(LAN) or Internet port.
2.
Access Ethernet configuration through: SETUP>MORE>COMM>INET.
3.
Follow the setup sequence as shown in the illustrations that follow, and edit
the Instrument and Gateway IP addresses and Subnet Mask to the desired
settings. (From the computer, enter the same information through an
application such as HyperTerminal).
the Instrument and Gateway IP addresses and Subnet Mask to the desired
settings. (From the computer, enter the same information through an
application such as HyperTerminal).
ENTR accept
new settings
EXIT ignores
new settings
SETUP X.X
PRIMARY SETUP MENU
CFG DAS
RNGE
PASS CLK
MORE
EXIT
SAMPLE
ENTER SETUP PASS : 818
8 1 8
ENTR
EXIT
SAMPLE
CO2 RNG=20.00 % CO2=XXX.XX
< TST TST > CAL
SETUP
SETUP X.X
COMMUNICATIONS MENU
ID
INET
COM1
COM2
EXIT
SETUP X.X
SECONDARY SETUP MENU
COMM
VARS DIAG
EXIT
SETUP X.X
DHCP: ON
ON
ENTR EXIT
SETUP X.X
DHCP: ON
<SET SET>
EDIT
EXIT
SETUP X.X
DHCP: OFF
OFF
ENTR
EXIT
(continues in next illustration)
07276B DCN6418