Teledyne API T803 CO2/O2 Analyzer Operation Manual 
Data Acquisition System (DAS & APICOM 
 
 145 
7.1.5.8. ENABLING / DISABLING THE HOLD OFF FEATURE 
The DAS  HOLD OFF feature prevents data collection during calibration 
operations.   
To enable or disable the HOLD OFF, follow the instruction shown in Section 
7.1.5 then press: 
Starting at the EDIT CHANNEL MENU
SETUP X.X       
0) CONC:  ATIMER 2, 4032, RS232
PREV NEXT
INS
DEL
EDIT
PRNT
EXIT
SETUP X.X       
 NAME: CONC
SET>
EDIT
EXIT
Continue pressing <SET or SET> until ...
SETUP X.X       
CAL.HOLD OFF: OFF
<SET SET
EDIT
EXIT
Press PREV and 
NEXT to scroll to the
DATA CHANNEL to 
be edited
SETUP X.X       
CAL.HOLD OFF: OFF
OFF
ENTR EXIT
EXIT discards the new 
setting
ENTR accepts the 
new setting
Toggle to turn the 
HOLDOFF feature 
ON/OFF
 
 
HOLD OFF also prevents DAS measurements from being made at certain times 
when the quality of the analyzer’s O
2
 measurements may be suspect (e.g. while 
the instrument is warming up).  In this case, the length of time that the HOLD 
OFF 
feature is active, is determined by the value of the internal variable (VARS), 
DAS_HOLD OFF.   
To set the length of the DAS_HOLD OFF period, see Section 11.1.3. 