Data Acquisition System (DAS & APICOM 
Teledyne API T803 CO2/O2 Analyzer Operation Manual 
146  
 
7.1.5.9. THE COMPACT REPORT FEATURE 
When enabled, this option avoids unnecessary line breaks on all RS-232 reports.  
Instead of reporting each parameter in one channel on a separate line, up to five 
parameters are reported in one line.   
The  COMPACT DATA REPORT generally cannot be accessed from the 
standard DAS front panel menu, but is available via the instrument’s 
communication ports by using APICOM or the analyzer’s standard serial data 
protocol. 
7.1.5.10. THE STARTING DATE FEATURE 
This option allows specifying a starting date for any given channel when the user 
wants to start data acquisition only after a certain time and date.  If the 
STARTING DATE
 is in the past (the default condition), the DAS ignores this 
setting and begins recording data as defined by the REPORT PERIOD setting. 
The  STARTING DATE generally cannot be accessed from the standard DAS 
front panel menu, but is available via the instrument’s communication ports by 
using APICOM or the analyzer’s standard serial data protocol. 
07276B DCN6418