Teledyne Tablet Accessory T803 Manual De Usuario

Descargar
Página de 304
Teledyne API T803 CO2/O2 Analyzer Operation Manual 
Troubleshooting and Service 
 
 
189 
11.2.2.2. RELAY PCA STATUS LED S 
 
D6 (Yellow) O
2
 Sensor Heater 
D5 (Yellow) – CO
2
 Sensor Heater 
(only with CO
2
 option)
 
D1 (RED) 
Watchdog Indicator
 
Figure 11-4: 
Relay PCA Status LEDs Used for Troubleshooting 
 
Table 11-4:  Relay PCA Status LED Failure Indications 
LED Color 
Function 
 Status When LED Lit 
(Energized State) 
 Status When LED Unlit 
(Default State)
 
D1 
Red Watchdog 
Circuit 
Cycles ON/OFF every 3 Seconds  
under direct control of the analyzer’s CPU. 
D2-D4 
SPARE 
D5
 
Yellow CO
2
  Sensor Cell heater 
Heating 
Not Heating 
D6
 
Yellow O
2
 Sensor heater 
Heating 
Not Heating 
D7
2
 
Green 
 
 
 
D8
Green 
 
 
 
D9
2
 Green 
 
 
 
D10
Green 
 
 
 
D11 - 16 
SPARE 
 Not Used 
07276B DCN6418