Troubleshooting and Service 
Teledyne API T803 CO2/O2 Analyzer Operation Manual 
190 
 
11.3. GAS FLOW PROBLEMS 
When troubleshooting flow problems, it is a good idea to first confirm that the 
actual flow and not the analyzer’s flow sensor and software are in error, or the 
flow meter is in error.  Use an independent flow meter to perform a flow check as 
described in Section 10.3.4.  If this test shows the flow to be correct, check the 
pressure sensors as described in Section 11.6.8. 
In general, flow problems can be divided into three categories: 
  Flow is too high 
  Flow is greater than zero, but is too low, and/or unstable 
  Flow is zero (no flow) 
When troubleshooting flow problems, it is crucial to confirm the actual flow rate 
without relying on the analyzer’s flow display.  The use of an independent, 
external flow meter to perform a flow check as described in Section 10.3.4 is 
essential.  
The flow diagrams found in a variety of locations within this manual depicting 
the T803 in its standard configuration and with options installed, can help in 
trouble-shooting flow problems.  For your convenience the diagrams are collected 
here. 
11.3.1. T803 INTERNAL GAS FLOW DIAGRAMS 
 
 
Figure 11-5: 
T803 – Internal Gas Flow 
07276B DCN6418