Teledyne Tablet Accessory T803 Manual De Usuario

Descargar
Página de 304
Troubleshooting and Service 
Teledyne API T803 CO2/O2 Analyzer Operation Manual 
192 
 
11.3.2.4. DISPLAYED FLOW WARNINGS 
This warning means that there is inadequate gas flow.  There are four conditions 
that might cause this: 
  A leak upstream or downstream of the flow sensor 
  A flow obstruction upstream or downstream of the flow sensor 
  Bad Flow Sensor Board 
 Bad 
pump 
To determine which condition is causing the flow problem, view the sample 
pressure and sample flow functions on the front panel display.  If the sample 
pressure is reading abnormally low, then the cause is likely a flow obstruction 
upstream of the flow sensor.  First, check the sample filter and make sure it is not 
plugged and then systematically check all the other components upstream of the 
orifice to ensure that they are not obstructed.   
If the sample pressure is reading normal but the sample flow is reading low, then 
it is likely that the pump diaphragm is worn or there is an obstruction downstream 
of the flow sensor. 
11.3.2.5. ACTUAL FLOW DOES NOT MATCH DISPLAYED FLOW 
If the actual flow measured does not match the displayed flow, but is within the 
limits of 110-130 cm
3
/min, adjust the calibration of the flow measurement as 
described in Section 10.3.4. 
11.3.2.6. SAMPLE PUMP  
The sample pump should start immediately after the front panel power switch is 
turned ON.  With the SAMPLE inlet plugged the test function PRES should read 
less than 10”-Hg for a pump in good condition.  Readings above 10” Hg indicate 
that the pump needs rebuilding.  If the test function SAMP FL is greater than 10 
cm
3
/min there is a leak in the pneumatic lines. 
11.4. CALIBRATION PROBLEMS 
11.4.1. MISCALIBRATED 
There are several symptoms that can be caused by the analyzer being 
miscalibrated.  Miscalibration is indicated by out-of-range Slopes and Offsets as 
displayed through the test functions and is frequently caused by the following: 
  Bad span gas:  This can cause a large error in the slope and a small error in 
the offset.  Delivered from the factory, the T803’s slope is within ±15% of 
nominal.  Bad span gas will cause the analyzer to be calibrated to the wrong 
value.  If in doubt have the span gas checked by and independent lab. 
  Contaminated zero gas:  Excess H2O can cause a positive or negative offset 
and will indirectly affect the slope. 
  Dilution calibrator not set up correctly or is malfunctioning:  This will also 
cause the slope, but not the zero, to be incorrect.  Again the analyzer is being 
calibrated to the wrong value. 
  Too many analyzers on the manifold:  This can cause either a slope or offset 
error because ambient gas with its pollutants will dilute the zero or span gas.   
07276B DCN6418