Principles of Operation 
Teledyne API T803 CO2/O2 Analyzer Operation Manual 
206 
 
 
Figure 12-1: 
Paramagnetic O
2
 Sensor Design 
When sample gas containing oxygen flows into the magneto-mechanical sensor, 
the O2 molecules are drawn toward the strong point of the magnetic field. This 
causes the N2 filled spheres to deflect even more so that the suspended dumbbell 
body pivots on the platinum wire. The more O2 present the further the dumbbell 
body is deflected from its neutral position. 
The position of the dumbbell is detected by a pair of photocells that receive a 
light beam reflected from a mirror attached to the center of the dumbbell body. 
As the dumbbell body pivots, the angle of the reflected light beam on the 
photocells changes. The resulting potential difference creates a current. 
Light 
Source
Photocells
Coil
 
Figure 12-2: 
Paramagnetic O
2
 Sensor Block Diagram 