Principles of Operation 
Teledyne API T803 CO2/O2 Analyzer Operation Manual 
208 
 
chamber.  This dual wavelength method of measuring CO
2
 allows the instrument 
to compensate for ancillary effects like sensor aging and contamination. 
12.3. OPERATION WITHIN THE T803 ANALYZER 
Operationally, both the CO
2
 and O
2
 sensors are transparently integrated into the 
core analyzer operation.  All functions can be viewed or accessed through the 
front panel. 
 The 
CO
2
 concentration is displayed in the upper right-hand corner, alternating 
with O
2
 concentration.   
  Test functions for the slope and offset of CO
2
 and of O
2
 are viewable from the 
front panel along with the analyzer’s other test functions. 
  Calibration of each 
 
sensor is performed via the front panel CAL.  See 
Section 9 for more details
Stability of each
 
sensor can be viewed via the front panel (see Section 9).   
 
 
12.3.1.1. ELECTRONIC OPERATION OF THE CO
2
 SENSOR 
The CO
2
 PCA is powered by 12 VDC from the analyzer via the relay card, which 
outputs a 0-5 VDC analog signal to the analyzer’s CPU via the motherboard that 
corresponds to the concentration of CO
2
 measured by the probe. 
 
Figure 12-4. 
CO
2
 Sensor PCA Layout and Electronic Connections 
 
07276B DCN6418