Principles of Operation 
 
215 
12.6.3. RELAY BOARD 
The CPU issues commands via a series of relays and switches located on a 
separate printed circuit assembly, called the relay PCA, to control the function of 
key electromechanical devices such as heaters.  The relay PCA receives 
instructions in the form of digital signals over the I
2
C bus, interprets these digital 
instructions and activates its various switches and relays appropriately. 
The relay PCA is located in the right-rear quadrant of the analyzer and is 
mounted vertically on the backside of the same bracket as the instrument’s DC 
power supplies.   
 
Power 
Connection 
for DC 
Heaters 
Status LED’s 
(D2 through D16) 
DC Power Supply 
Test Points 
Watchdog 
Status LED (D1) 
(JP5) 
Thermocouple 
Configuration 
Jumpers  
Thermocouple 
Signal Output 
I
2
C
 Connector
Valve Control 
Drivers 
Pump Power  
Output 
(JP7) 
Pump AC   
Configuration 
Jumper 
AC Power  
IN 
DC Power 
Distribution 
Connectors 
Valve Control 
Connector 
(J2) 
Connector for 
AC Relays  
K4 & K5 
(J18)
Connector for AC Relays K4 & K5
Heater AC Power 
Configuration 
Jumpers 
JP2
JP
 
 
 
 
Figure 12-9: 
Relay PCA Layout (PN 04135) 
 
 
CAUTION 
E
LECTRICAL 
S
HOCK 
H
AZARD
 
Only those relays actually required by the configuration of the T803 are populated. 
A protective retainer plate is installed over the AC power relays to keep them securely 
seated in their sockets and prevent accidental contact with those sockets that are not 
populated see Figure 12-10)

Never remove this retainer while the instrument is plugged in and turned on.  The contacts of 
the AC relay sockets beneath the shield carry high AC voltages even when no relays are 
present. 