Principles of Operation 
Teledyne API T803 CO2/O2 Analyzer Operation Manual 
216 
 
 
 
AC Relay Retainer 
Plate 
Retainer  
Mounting  
Screws 
 
Figure 12-10:  Relay PCA with AC Relay Retainer in Place 
12.6.3.1. STATUS LEDS  
LEDs located on the Analyzer’s relay PCA, show the current status of various 
control functions performed by the relay PCA  The three that are used in the T803 
are described in Table 12-1, and their locations are illustrated in Figure 12-11. 
Table 12-1:  Relay PCA Status LEDs 
LED Color 
Function 
Status When LED Lit 
(Energized State) 
Status When LED Unlit 
(Default State)
 
D1 
Red Watchdog 
Circuit 
Cycles ON/OFF every 3 Seconds  
under direct control of the analyzer’s CPU. 
D2-D4 
SPARE 
D5
 
Yellow CO
2
  Sensor Cell heater 
Heating 
Not Heating 
D6
 
Yellow O
2
 Sensor heater 
Heating 
Not Heating 
D7
2
 
Green 
 
 
 
D8
Green 
 
 
 
D9
2
 Green 
 
 
 
D10
Green 
 
 
 
D11 - 16 
SPARE 
2
  Not Used. 
07276B DCN6418