Teledyne API T803 CO2/O2 Analyzer Operation Manual 
Getting Started 
 
 51 
after the 60-minute warm up period is over, investigate their cause using the 
troubleshooting guidelines in Section 11.   
Table 3-9:  Warning Messages  
MESSAGE DEFINITION 
ANALOG CAL WARNING 
The instrument's A/D circuitry  or one of its analog outputs is not calibrated. 
BOX TEMP WARNING 
The temperature inside the T803 chassis is outside the specified limits. 
CANNOT DYN SPAN
2
 
Remote span calibration failed while the dynamic span feature was set to ON 
CANNOT DYN ZERO
3
 
Remote zero calibration failed while the dynamic zero feature was set to ON 
CONFIG INITIALIZED 
Configuration was reset to factory defaults or was erased. 
CO2  ALRM1 WARNING
4 
CO
2
 concentration alarm limit 1 exceeded 
CO2  ALRM2 WARNING
4 
CO
2
 concentration alarm limit 2 exceeded 
CO2 CELL TEMP WARN 
CO
2
 sensor cell temperature outside of warning limits specified by 
CO2_CELL_SET
 variable. 
DATA INITIALIZED 
DAS data storage was erased. 
O2 ALRM1 WARNING
4 
O
2
 concentration alarm limit 1 exceeded 
O2 ALRM2 WARNING
4 
O
2
 concentration alarm limit 2 exceeded 
O2 CELL TEMP WARN 
O
2
 sensor cell temperature outside of warning limits specified by 
O2_CELL_SET
 variable. 
REAR BOARD NOT DET 
CPU unable to communicate with the motherboard. 
RELAY BOARD WARN 
CPU is unable to communicate with the relay PCA. 
SAMPLE FLOW WARN 
The flow rate of the sample gas is outside the specified limits. 
SAMPLE PRESS WARN 
Sample gas pressure outside of operational parameters. 
SYSTEM RESET
1
 
The analyzer was rebooted or the CPU was reset. 
1
  Does not clear after power up. 
2
  Clears the next time successful zero calibration is performed. 
3
  Clears the next time successful span calibration is performed. 
Only active if the Concentration Alarm Option is installed  
 
07276B DCN6418