Teledyne API T803 CO2/O2 Analyzer Operation Manual 
Getting Started 
 
 49 
3.3.2.9. EXHAUST OUTLET   
Attach an exhaust line to the analyzer’s EXHAUST outlet fitting.  The exhaust 
line should be: 
  PTEF tubing; minimum O.D ¼”  
  A maximum of 10 meters long. 
  Vented outside the T803 analyzer’s enclosure, preferably outside the shelter 
or at least into a well-ventilated area. 
 
Figure 3-17: 
T803 Internal Gas Flow 
I
MPORTANT
 
Leak Check: 
     Run a leak check once the appropriate pneumatic connections 
have been made; check all pneumatic fittings for leaks using the 
procedures defined in Section 10.3.3. 
 