Teledyne Tablet Accessory T803 Manual De Usuario

Descargar
Página de 304
Setup Menu 
Teledyne API T803 CO2/O2 Analyzer Operation Manual 
78    
Table 5-1:  Variable Names (VARS) 
NO. VARIABLE 
DESCRIPTION 
ALLOWED 
SETTINGS
VARS
 
Default 
settings
 
DAS_HOLD_OFF 
Changes the Internal Data Acquisition System (DAS) HOLD 
OFF
 timer: 
No data are stored in the DAS channels during situations 
when the software considers the data to be questionable 
such as during warm up or just after the instrument returns 
from one of its calibration modes to SAMPLE Mode.   
May be set for 
intervals 
between  
0.5 – 20 min 
 
15 min. 
STABIL_GAS
 
Selects which gas measurement is displayed when the 
STABIL test function is selected. 
CO
2
; O
2
 
CO2 
TPC_ENABLE 
NOTE: It is strongly recommended that this variable 
NOT be changed. 
ON enables, OFF disables temperature and pressure 
compensation  
ON, OFF 
ON 
DYN_ZERO 
Special configuration; 
consult factory 
[Automatically adjusts offset and slope of the O
2
 response 
when performing a zero point calibration during an AutoCal.] 
[ON, OFF] 
[OFF] 
DYN_SPAN 
Special configuration; 
consult factory
 
[Automatically adjusts offset and slope of the O
2
 response 
when performing a zero point calibration during an AutoCal.] 
[ON, OFF] 
[OFF] 
CONC_PRECISION 
Allows the user to set the number of significant digits to the 
right of the decimal point display of concentration and 
stability values. 
AUTO, 1, 2, 
3, 4 
AUTO 
CLOCK_ADJ
 
Adjusts the speed of the analyzer’s clock.  Choose the + 
sign if the clock is too slow, choose the - sign if the clock is 
too fast. Toggle buttons to change the number of seconds. 
-60 to +60 
s/day 
0 Sec/Day
SERVICE_CLEAR 
Pressing the OFF button to display SERVICE_CLEAR:ON, 
followed by pressing ENTR resets the service interval timer 
and returns this Var back to its default setting, ready for the 
next reset. 
ON, OFF
 
OFF 
TIME_SINCE_SVC 
Displays time in hours since last service (restarted by the 
SERVICE_CLEAR Variable). 
0-50,000 
SVC_INTERVAL 
Sets the interval in hours between service reminders. 
0-100,000 
 
07276B DCN6418