Teledyne Tablet Accessory T803 Manual De Usuario

Descargar
Página de 304
Teledyne API T803 CO2/O2 Analyzer Operation Manual 
Setup Menu 
 
 79 
To access and navigate the VARS menu, use the following button sequence. 
SETUP X.X         6) CLOCK_ADJUST=0 Sec/Day
 PREV
NEXT
JUMP
EDIT
ENTR EXIT
SETUP X.X             CLOCK_ADJUST=0 Sec/Day
+
0
ENTR EXIT
Enter sign and number of 
seconds per day the clock 
gains (-) or loses(+)
SAMPLE
CO2 RNG=20.00 %            CO2=XXX.XX
<TST
TST> CAL
                  MSG
SETUP
SETUP X.X          
0) DAS_HOLD_OFF=15.0 Minutes
PREV
NEXT
JUMP
EDIT
PRNT EXIT
SETUP X.X       
 PRIMARY SETUP MENU
 CFG ACAL DAS
RNGE
PASS CLK
MORE
EXIT
SETUP X.X         
SECONDARY SETUP MENU
 COMM
VARS
DIAG
EXIT
SETUP X.X    
ENTER PASSWORD:818
 8
  1
  8
ENTR
EXIT
Toggle to enter 
the correct
PASSWORD
In all cases:
EXIT discards the new 
setting
ENTR accepts the 
new setting
SETUP X.X         3) DYN_ZERO=OFF
PREV
NEXT
JUMP
EDIT PRNT EXIT
SETUP X.X          4) DYN_SPAN=OFF
PREV
NEXT
JUMP
EDIT PRNT EXIT
SETUP X.X            
DAS_HOLD_OFF=15.0 Minutes
1
5
.0
ENTR EXIT
Toggle these keys to set 
the iDAS HOLDOFF time 
period in minutes 
(MAX =  20 minutes).
SETUP X.X          
5) CONC_PRECISION=AUTO
PREV
NEXT
JUMP
EDIT
PRNT EXIT
SETUP X.X           
CONC_PRECISION=AUTO
AUTO
1
2
3
4
ENTR EXIT
Press to select the 
precision of the gas 
concentration display
SETUP X.X          
1) STABIL_GAS=CO2
PREV
NEXT
JUMP
EDIT
PRNT EXIT
SETUP X.X           
STABIL_GAS=O2
O2
CO2
ENTR EXIT
SETUP X.X          
2) TPC_ENABLE=ON
PREV
NEXT
JUMP
EDIT
PRNT EXIT
SETUP X.X           
TPC_ENABLE:ON
ON
ENTR EXIT
Toggle to turn ON or OFF 
temperature pressure 
compensation. 
Press to select which gas 
will be reported by the 
STABIL test function 
 
07276B DCN6418