Teledyne API T803 CO2/O2 Analyzer Operation Manual
Setup Menu
79
To access and navigate the VARS menu, use the following button sequence.
SETUP X.X 6) CLOCK_ADJUST=0 Sec/Day
PREV
NEXT
JUMP
EDIT
ENTR EXIT
SETUP X.X CLOCK_ADJUST=0 Sec/Day
+
0
0
ENTR EXIT
Enter sign and number of
seconds per day the clock
gains (-) or loses(+)
SAMPLE
CO2 RNG=20.00 % CO2=XXX.XX
<TST
TST> CAL
MSG
SETUP
SETUP X.X
0) DAS_HOLD_OFF=15.0 Minutes
PREV
NEXT
JUMP
EDIT
PRNT EXIT
SETUP X.X
PRIMARY SETUP MENU
CFG ACAL DAS
RNGE
PASS CLK
MORE
EXIT
SETUP X.X
SECONDARY SETUP MENU
COMM
VARS
DIAG
EXIT
SETUP X.X
ENTER PASSWORD:818
8
1
8
ENTR
EXIT
Toggle to enter
the correct
PASSWORD
In all cases:
EXIT discards the new
setting
ENTR accepts the
new setting
SETUP X.X 3) DYN_ZERO=OFF
PREV
NEXT
JUMP
EDIT PRNT EXIT
SETUP X.X 4) DYN_SPAN=OFF
PREV
NEXT
JUMP
EDIT PRNT EXIT
SETUP X.X
DAS_HOLD_OFF=15.0 Minutes
1
5
.0
ENTR EXIT
Toggle these keys to set
the iDAS HOLDOFF time
period in minutes
(MAX = 20 minutes).
SETUP X.X
5) CONC_PRECISION=AUTO
PREV
NEXT
JUMP
EDIT
PRNT EXIT
SETUP X.X
CONC_PRECISION=AUTO
AUTO
1
2
3
4
ENTR EXIT
Press to select the
precision of the gas
concentration display
SETUP X.X
1) STABIL_GAS=CO2
PREV
NEXT
JUMP
EDIT
PRNT EXIT
SETUP X.X
STABIL_GAS=O2
O2
CO2
ENTR EXIT
SETUP X.X
2) TPC_ENABLE=ON
PREV
NEXT
JUMP
EDIT
PRNT EXIT
SETUP X.X
TPC_ENABLE:ON
ON
ENTR EXIT
Toggle to turn ON or OFF
temperature pressure
compensation.
Press to select which gas
will be reported by the
STABIL test function
07276B DCN6418