Waveform Programming Language
3-188
AWG710&AWG710B Arbitrary Waveform Generator User Manual
In a waveform expression, the data length of the file created and the clock 
information are determined as follows.
If <output–signal–name> is a marker:
If the output file does not already exist, an error will occur. Attributes such as 
the data length (output file size) and clock information are unchanged. The 
analog data section does not change. Neither the size or the close variable value 
is used. If the waveform expression includes a <signal–name>, then a file 
shorter than the output file would cause an error. If the input file is longer in 
this case, the data around the tail will not be used.
If <output–signal–name> is analog data:
A new file is always created without using the output file. Since the same file 
name may be specified for the input, the new file is tentatively created under 
another file name; then renamed.
All output file marker values will be 0. The output file data length and clock 
information will be as indicated on the screen.
If the waveform expression includes one or more <signal–name>s:
The output waveform length will equal to that of the shortest waveforms 
included in the <waveform–expression>. The clock information will match the 
one appearing first (that is, the one coded at the leftmost) out of those used in 
the waveform expression.
If the waveform expression includes no <signal–name>:
The output waveform length depends on the size variable value. The clock 
value depends on the clock variable value.