Function Reference — SCAN_Sequence_Setup
©
 National Instruments Corporation
2-305
NI-DAQ FRM for PC Compatibles
gainVector contains the gain settings to be used for each channel in chanVector. The channel 
listed in entry of chanVector will have the gain listed in entry i of gainVector.
scanRateDivisorVector contains the scan rate divisors to be used for each channel. The 
sample rate for a channel equals the base scan rate (that is, the scan rate specified when 
SCAN_Start
 is called) divided by the scan rate divisor for that channel. The channel listed in 
entry of chanVector will have the scan rate divisor listed in entry of 
scanRateDivisorVector.
scansPerSequence is an output parameter that contains the total number of scans in the scan 
sequence created by NI-DAQ from your chanVector and scanRateDivisorVector including 
any scans that consist entirely of ghost channels, or place holders.
samplesPerSequence is an output parameter that contains the total number of samples in the 
scan sequence excluding any ghost channel place holders. The total size of a scan sequence 
including ghost channel place holders is limited by the size of the memory on your device 
used to hold this information. Currently, this limit is 512 entries. Because 
samplesPerSequence excludes ghost channel place holders, an error might result even if 
samplesPerSequence is less than 512.
Using This Function
You must observe the following restrictions:
Interval scanning must be used.
A channel can be listed only once in the channel vector.
SCXI cannot be used.
The AMUX-64T device cannot be used.
Your acquisition cannot be pretriggered.
The size of your buffer (the value of the count parameter to 
SCAN_Start
) must be a 
multiple of samplesPerSequence.
The following example shows how to use 
SCAN_Sequence_Setup
:
numChans = 3
chanVector = {2, 5, 7}
gainVector = {1, 1, 1}
scanRateDivisorVector = {1, 2, 4}
The scan rate divisor for channel 2 is 1 so it will be sampled at the base scan rate. The scan 
rate divisor for channel 5 is 2 so it will be sampled at a rate equal to the base scan rate divided 
by 2. Likewise, the scan rate divisor for channel 7 is 4 so it will be sampled at a rate equal to 
the base scan rate divided by 4.