Chapter 2
Function Reference — WFM_from_Disk
©
 National Instruments Corporation
2-419
NI-DAQ FRM for PC Compatibles
If you have changed the analog output configuration from the defaults by changing the 
jumpers on the device, you must call 
AO_Configure
 to set the software copies of the 
settings prior to calling 
WFM_from_Disk
.
NI-DAQ ignores the group settings made by calling 
WFM_Group_Setup
 when you call 
WFM_from_Disk
WFM_from_Disk
 and the settings are not changed after you execute 
WFM_from_Disk
.
Note
For the MIO-16 and AT-MIO-16D, counter 2 must be available in order to use 
waveform generation. If an interval scan is in progress (see
SCAN_Start
) or a 
CTR function is using counter 2, waveform generation cannot proceed.
For the AT-MIO-16F-5, AT-MIO-64F-5, and AT-MIO-16X, you can use counters 1, 2, and 5, 
as well as a dedicated external update signal, to generate either interrupts or DMA requests. 
If you use counter 1 or 2, a RTSI line must also be available. NI-DAQ uses the first available 
counter among counters 5, 2, and 1, in that order.
For Lab and 1200 Series analog output devices, if the rate is smaller than 15.26 pts/s and 
counter B0 is busy in a data acquisition or counting operation, waveform generation cannot 
proceed.
On Am9513-based devices, to externally trigger a waveform generation operation, you can 
do so by first changing the gating mode of the counter NI-DAQ uses.
WFM_from_Disk
 uses either the default gating mode (none) or the gating mode you specify 
through the 
CTR_Config
 function. You need to connect your trigger signal to the gate pin on 
the I/O connector. Refer to the 
CTR_Config
 function description for details.
On a variety of E Series devices, you can externally trigger a waveform generation in a variety 
of ways. Refer to the 
Select_Signal
 function for more details.