SCPI Command Reference
This section describes the Standard Commands for Programmable
Instruments (SCPI) commands for the Microwave Switch modules.
Commands are listed alphabetically by subsystem and also within each
subsystem.  
ABORt
The 
ABORt
 subsystem stops a scan in progress when the scan is enabled via
the interface, and the trigger modes are 
TRIGger:SOURce BUS
 or
TRIGger:SOURce HOLD
Subsystem Syntax 
ABORt
Comments
ABORt
 Operation invalidates the current channel list and sets
ARM:COUNt 1
 (one scanning cycle per 
INITiate
 command), sets
INITiate CONTinuous OFF
 (no continuous scanning cycles), and sets
TRIGger:SOURce IMMediate
 (continuous internal triggering).
Stopping Scans Enabled from Interface: When a scan is enabled
from the interface, use an interface 
CLEAR
 command or the HP
E1301 front panel "Reset Instr" or "Clear Instr" key to stop the scan.
When the scan is enabled from the interface and the trigger source is
TRIGger:SOURce BUS
 or 
TRIGger:SOURce HOLD
, use 
ABORt
 or
the HP E1301 front panel "Reset Instr" or "Clear Instr" keys to stop
the scan.
Stopping Scans Enabled from Front Panel: When a scan is
enabled from the HP E1301 front panel, execute 
*RST
 over the
interface or the front panel "Reset Instr" or "Clear Instr" keys to stop
the scan.
Related Commands: 
ARM
INITiate:CONTinuous
[ROUTe:]SCAN
,
TRIGger
Example
Stopping a Scan with 
ABORt
This example stops a (continuous) scan in progress.
TRIG:SOUR BUS
BUS is trigger source.
INIT:CONT ON
Set continuous scanning.
SCAN   (@100:102)
Scan channels 00-02.
INIT
Starts scanning cycle, close
channel 00.
ABOR
Aborts scan in progress.
50   HP E1368A/69A/70A Command Reference