DISPlay
The 
DISPlay
 subsystem monitors the channel state of a selected module (or
card) in a switchbox.  This subsystem only operates with mainframes that
have a display, such as the HP 75000 Series B Mainframe (HP Model
Number E1301A). 
Subsystem Syntax
DISPlay
:MONitor
 :CARD 
<
number
 | AUTO>
[:STATe] <
mode
[:STATe]?
:MONitor:CARD
DISPlay:MONitor:CARD <
number | 
AUTO>
  selects the module in a
switchbox to be monitored when the monitor mode is enabled.  Use the
DISPlay:MONitor:STATe
 command to enable and disable the monitor mode.
Parameters
Parameter Name
Parameter Type
Range of Values
Default Value
number
 | 
AUTO
numeric
1 - 99 or AUTO
AUTO
Comments
Selecting a Specific Module to be Monitored: Use the
DISPlay:MONitor:CARD
 command to send the card number in a
switchbox to be monitored.
Selecting the Present Module to be Monitored: Use the
DISPlay:MONitor AUTO
 command to select the last module
addressed by a switching command (e.g.,
[ROUTe:]CLOSe
).
*RST Condition:
 DISPlay:MONitor:CARD AUTO
Example
Select Module #2 in a Switchbox for Monitoring
DISP:MON:CARD 2
Selects module #2 in a switchbox
to be monitored when the mode is
enabled.
Chapter 4