(2) How to Inject Oil   
 
Inject oil through the oil injection port so that oil level reaches the center of the red circle in the 
oil gauge.   
 
 
 
 
 
 
 
 
 
 
 
 
 
  NOTICE:  If the oil level is below the lowest point of the red circle, pump performance may 
be degraded or oil may be deteriorated earlier than usual.    Conversely, if the amount of oil is 
excessive,  
oil may be blown out.    Hence, the proper level of oil should be maintained. 
 
(3) The oil used in this pump is NEOVAC MR-100 (supplied by Matsumura Sekiyu). 
 
(4) If oil has accumulated in the oil holder located at the center bottom of the rotary pump, you 
should push up the oil holder from the bottom of the base, remove the oil holder carefully and 
dispose of the oil in the oil holder.   
 
 
 
 
 
 
 
 
4.2.2    Replacement of Oil Mist Trap
   
 
The oil mist trap (L) is of a disposable type.    If oil splashes or oil fumes are generated from the 
oil mist trap, replace the trap with a new one (P/N 589-8652).   
 
Pull out
Press
Oil holder
Base
Oil 
gauge
Oil injection plug 
(injection port)
Proper oil 
level range 