4.3    Maintenance of Air CompressorOnly for P/N 539-9010 and 539-9060 
 
NOTICE 
Instruments shipped to Europe area where CE marking conformity is necessary do not provide 
with the air compressor. Also instruments those shall be compliant to SEMI-S2 guideline do not 
provide with the air compressor even when shipped to outside of Europe area. 
 
4.3.1  Checkup and Maintenance 
 
Check the following items regularly. 
Before checkup/maintenance, be sure turn the main switch off.   
 
Service Cycle 
Check Item 
Procedure 
Every 
day 
Every 
250
 h 
(1 month) 
Every 
3,000
 h 
(1 year) 
Every 
6,000
 h 
(2 years) 
Discharge of drain 
Loosen the drain 
cock of air tank to 
discharge drain. 
 
 
 
 
Activation of control 
devices (pressure 
gauge, pressure switch 
and safety valve) 
Check the 
working pressure 
and activated 
function. 
 
 
 
 
Abnormal vibration and 
abnormal sound 
 
 
 
 
 
Loose bolts, nuts and 
screws 
If loose, retighten 
them. 
 
 
 
 
Contamination and 
clogging of strainer/filter 
Blow air to 
remove clogged 
substance. 
 
 
 
 
Leakage through air 
valve 
 
 
 
 
 
● 
Piston ring 
 
 
 
 
 
 
Rider ring 
 
 
 
 
 
 
Ball bearing 
 
 
 
 
 
Bearing
 
Needle roller 
bearing 
 
 
 
 
 
 
 CAUTION 
The air compressor should be drained gradually by slowly loosening the drain cocks. 
The inside of tank is at a high pressure.    So, refrain from an abrupt cock operation, 
otherwise moisture may spray out to cause a danger.