Hitachi Microscope & Magnifier S-4800 Manual De Usuario

Descargar
Página de 294
 
4-
12
4.4  Troubleshooting 
 
4.4.1    When Column Evacuation does not Work 
 
When the column evacuation system does not work, check the followings. 
 
(1) Is an error message shown?     
If an error message is shown on the evacuation control panel, refer to XXXXXX to find the 
cause of the error and make countermeasure to it. 
 
(2) Check the TMP lamps on the evacuation control panel. Both NORMAL and POWER lamp 
shall be lit in normal evacuation condition. The POWER lamp will be lit about 10 minutes after 
starting evacuation. The Normal lamp will be lit further after about 10 minutes. 
   
(3) If you do not find any failure mentioned above, or could not make effective countermeasure, 
contact a service engineer. 
 
 
4.4.2    When Specimen Exchange Chamber Vacuum is not Good 
 
When about 10 minutes has been passed after starting SEC evacuation by EVAC switch or Open 
switch and the vacuum of the specimen exchange chamber does not reach to the specified value, 
buzzer will sound and the error E-17 will be shown on the evacuation control panel. Check the 
follows. 
(If the specimen is wet and therefore needs more evacuation time, leave it. Buzzer and error 
indication will stop by STOP button. When vacuum reaches to specified value, evacuation 
sequence will continue to next step even after causing the E-17 error.) 
 
(1) Introduce air into the specimen exchange chamber by the AIR button. Clean up and grease 
up the O-ring seal of the chamber.    Then start evacuation again. 
 
(2)   Check if your specimen is not wet.    If the specimen is not dry, outgas from the specimen 
may cause long evacuation time.   
 
(3) Check oil level of rotary pumps.    If the level is not normal, add or exchange oil of rotary 
pumps referring to instruction manual. 
 
 CAUTION 
Rotary pump 
 
Beware of high 
temperature
 
The surface of the rotary pump reaches a temperature of 
about 70
°
C while the pump is operating.     
Touching it could result in a serious burn.