Hitachi Microscope & Magnifier S-4800 Manual De Usuario

Descargar
Página de 294
 
4-
22
(6) Mount the heater to the electron gun chamber. 
The heater consists of 2 parts. Fix them with the heater band.   
 
 
 
 
(7) Put two rods of inner bake read wire 
into the HV cable socket. Insert rods 
correctly to pins A and B at the bottom 
of the socket referring the right side 
picture. 
Be careful not to insert to other pins.   
 
(8) Remove the cover of baking power 
connectors provided at the right side of 
ion pump cover. 
   
 
 
(9) Connect the lead wire for OUTER 
BAKE and INNER BAKE to the 
connectors.   
Put the cover again. (A protection is 
provided so as if the cover is not put, 
baking will not start) 
 
(10) Attach the baking protection cover to 
the top part of the electron gun. It is for 
isolation from high temperature part.   
( A protection is provided so that 
baking will not start if the cover is not 
attached correctly)   
 
NOTICE 
Confirm that heater lead wire does not touch to the heater directly.     
 
 CAUTION 
Electron gun/ion pump baking 
 
Beware of high 
temperature
 
  The electron gun and ion pumps are heated at about 
250
°
C in order to keep the electron gun in a high 
vacuum.    Never touch these parts. Otherwise you will 
suffer a burn.