PRECAUTION - 12  
(4) Floor 
vibration 
S-4800 Type I 
 
 
 
 
 
 
 
 
 
 
 
 
 
S-4800 Type II 
 
 
 
 
 
 
 
 
 
 
 
 
 
The allowable amplitude corresponds to an image deviation of less than 0.2 mm on a 
micrograph at magnification 220,000
×
 
 
(5)  Acoustic noise .................. The following table shows the relation between allowable noise 
level (C characteristic) and frequency at the S-4800 installation 
site.    The allowable amplitude corresponds to an image 
deviation of less than 0.2 mm on a micrograph at magnification 
220,000
×
 
 
 
 
 
Oscillation 
Frequency 
(Hz) 
Allowable Amplitude (X, Y) 
 
(
µ
m Peak to Peak) 
1.0   
Less than 7.0 
1.4   
Less than 2.0 
2.0   
Less than 2.0 
3.0   
Less than 4.0 
4.0   
Less than 4.5 
5.0   
Less than 4.5 
6.0   
Less than 4.0 
10.0   
Less than 3.0 
Oscillation
Frequency
(Hz) 
Allowable Amplitude (Z) 
 
(
µ
m Peak to Peak) 
1.0   
Less than 25.0 
2.0   
Less than 14.0 
3.0   
Less than 5.0 
3.5   
Less than 2.9 
4.0   
Less than 1.8 
5.0   
Less than 1.0 
6.0   
Less than 1.1 
10.0   
Less than 1.2 
Oscillation 
Frequency 
(Hz) 
Allowable Amplitude (X, Y) 
 
(
µ
m Peak to Peak) 
1.0   
Less than 8.0 
1.5   
Less than 3.0 
2.0   
Less than 1.4 
2.4   
Less than 1.4 
3.0   
Less than 3.0 
5.0   
Less than 4.0 
6.0   
Less than 3.5 
10.0   
Less than 3.5 
Oscillation
Frequency
(Hz) 
Allowable Amplitude (Z) 
 
(
µ
m Peak to Peak) 
1.0   
Less than 20.0 
2.0   
Less than 10.0 
3.0   
Less than 5.0 
4.0   
Less than 2.7 
5.0   
Less than 1.8 
6.0   
Less than 1.0 
7.0   
Less than 1.0 
10.0   
Less than 1.0 
Frequency : f (Hz)  Allowable Noise LevelType I) Allowable Noise Level (Type II 
f≦150 
75 dB or less 
72 dB or less 
150
≦f≦800 
60 dB or less 
60 dB or less 
800
≦f≦2000 
54 dB or less 
52 dB or less