PRECAUTION - 10  
(6)  Signal processing modes ....................Automatic brightness control, 
Gamma control, 
Automatic focus, 
Automatic stigmator 
 
(7)  Automatic data display ........................Image number, accelerating voltage, magnification, 
micron bar, micron value, date/time and working 
distance can be printed on the film. 
 
(8)  Data entry ............................................Alphanumeric characters, numbers, and marks can 
be written on the image from the keyboard. 
 
(9)  Electrical image shift............................
±
12 
µ
m (WD = 8 mm) 
 
 
6.  Evacuation System 
 
(1)  System type.........................................Fully automatic pneumatic-valve system 
 
(2)  Ultimate vacuum levels........................Specimen chamber 
×
 10
-4
 Pa 
Electron gun chamber : 
 
IP-1 1 
×
 10
-7
 Pa or better 
IP-2 2 
×
 10
-6
 Pa or better 
IP-3 7 
×
 10
-5
 Pa or better 
 
(3)  Vacuum pumps....................................Electron optical system :  3 ion pumps 
Specimen chamber 
:  Turbo molecular pump   
1 oil rotary pump should be provided by user. 
 
(4)  Compressor .........................................Oil-less type compressor should be provided by user. 
 
 
7.  Protection Devices 
 
Warning devices ..........................................Power failure, Cooling-water interruption, Inadequate 
vacuum