Language Dictionary 71
Table 7-3. Agilent SAS Programming Parameters (in Fixed Mode)
Parameter
Agilent E4350B
Agilent E4351B
CURR[:LEV] MAX and
8.16 A
4.08 A
CURR[:LEV]:TRIG MAX
8.16 A
4.08 A
CURR:PROT[:LEV]: MAX
10 A
5 A
(Programming range is 0 to MAX)
*RST Current Value
*RST Current Value
0.096 A
0.048 A
*RST Current Protect Value
MAX for both models
OUTP:PROT:DEL
*RST Value
*RST Value
0 to 32.767 s (MAX) both models
200 ms both models
VOLT[:LEV] MAX and
61.50 V
123.0 V
VOLT[:LEV]:TRIG MAX
61.50 V
123.0 V
(Programming range is 0 to MAX)
*RST Voltage Value
*RST Voltage Value
0 V for both models
VOLT:PROT MAX
73.0 V
140.0 V
(Programming range is 0 to MAX)
*RST OVP value
*RST OVP value
73.0 V
140.0 V
Calibration Commands
Display Subsystem
This subsystem controls the state and output of the alphanumeric portion of the display.
DISP
Enables or disables the display. When disabled, the display characters are blank. The annunciators are not affected by this
command.
command.
Command Syntax
DISPlay[:WINDow][:STATe] <bool>
Parameters
0 | 1 | OFF | ON
*RST Value
ON
Examples
DISP ON DISPLAY:STATE ON
Query Syntax
DISPlay[:WINDow][STATe]?
Returned Parameters
<NR1> 0 or 1
Related Commands
DISP:MODE DISP:TEXT *RST
DISP:MODE
Switches the display between its normal metering mode and a mode in which it displays text sent by the user.
Command Syntax
DISPlay[:WINDow]:MODE <CRD>
Parameters
NORMal | TEXT
*RST Value
NORM
Examples
DISP:MODE NORM DISPLAY:MODE NORMAL
Query Syntax
DISPlay[:WINDow]:MODE?
Returned Parameters
<CRD> NORMAL or TEXT
Related Commands
DISP DISP:TEXT *RST