72    Language Dictionary
 DISP:TEXT
 
 
Sends character strings to the display when the display mode is TEXT.  The LCD display has the following character set:
 
 LCD Character Set
 
uppercase letters
 
A through Z (Case-sensitive entry)
 
digits
 
0 through 9
 
punctuation
 
_  |  “  $  < >  +  -  /  =  ?  .  :  ,
 
blank space
 
 
 
A display is capable of showing up to 12 characters.  However, the three punctuation characters do not count toward the 12-
character limit when they are preceded by an alphanumeric character.  When punctuation characters are included, then the
maximum number of characters (alphanumeric + punctuation) that can be displayed is 15.  If it exceeds the display capacity,
a message will be truncated to fit and no error message will be generated.  If any character in the message is not a member
of the above character set, the character will be displayed as a starburst (all 16 segments of the character are lit).
 
 
Note
IEEE Standard Digital Interface for Programmable Instrumentation requires that a string be enclosed in
either single (‘) or double (") quotes.
 
 
Command Syntax 
DISPlay[:WINDow]:TEXT [:DATA] <STR>
 
Parameters 
(See LCD character set)
 
*RST Value
‘      ‘
 
Examples
DISP:TEXT "DEFAULT_MODE"
 
Query Syntax
DISPlay[:WINDow]:TEXT?
 
Returned Parameters
<STR>      (Last programmed text string)
 
Related Commands
DISP    DISP:MODE    *RST
 
 
 
 
 
Measure Subsystem
 
 
Measurement commands measure the output voltage and current.
 
 
 MEAS:CURR?
 MEAS:VOLT?
 
 
These queries return the voltage and current measured at the Agilent SAS’s sense terminals.
 
 
Query Syntax
MEASure:CURRent[:DC]?
 
MEASure:VOLTage[:DC]?
 
Parameters
(None)
 
Default Suffix
A  for MEAS:CURR?
 
V  for MEAS:VOLT?
 
Examples
MEAS:CURR?
MEAS:VOLT?
MEASURE:VOLTAGE:DC? MV
 
Returned Parameters