HP Cell Phone Accessories E6380-90019 Manual De Usuario
82
Chapter 4
CDMA Measurements - Screens and Control Fields
Average Power Measurements
Average Power Measurements
CDMA Measurements
Average Power Measurements
Avg Pwr
Avg Pwr
displays the average power level of all signals at the RF
IN/OUT port. The units (watts, volts, or dBm) can be changed by
positioning the cursor in front of the units field and then pressing the
desired units key. (See
positioning the cursor in front of the units field and then pressing the
desired units key. (See
.) Refer to the
HP 8935 CDMA Base Station Tests Applications Guide for more
information about average power measurements.
information about average power measurements.
Screens on which this measurement is displayed.
• CDMA ANALYZER
• CDMA GENERATOR
ADC FS
ADC FS
is displayed when average power is measured. ADC FS
indicates how close the measured power level is to the maximum
allowed input level at the RF IN/OUT port for a particular
allowed input level at the RF IN/OUT port for a particular
Pwr Gain
setting (on the CDMA ANALYZER screen). Changes to ADC FS can be
made by altering the
made by altering the
Pwr Gain
field setting and by altering the signal
level into the RF IN/OUT port. ADC FS should be between
−
1.0 and
−10
dB when the
Pwr Gain
field is set to
Auto
, and can be manually set up
to
−
0.5 dB when set to
Hold
. The measurement will be aborted if ADC
FS goes above
−
0.5 dB. Measurement accuracy may be degraded if ADC
FS goes below
−
10 dB.
Screens on which this measurement is displayed.
• CDMA ANALYZER
• CDMA GENERATOR