Cornelius UF-1 Manual De Usuario

Descargar
Página de 36
UF-1 and UFB-1 Valves Training Manual
- 8 -
© 1999-2005, IMI Cornelius Inc.
2. S
YRUP
/C
ONCENTRATE
Syrup should always be precooled before setting the ratio. Syrup takes a path through the 
valve parallel to the water path. It is introduced in the block, travels through a syrup flow control, 
banjo, valve head and out the nozzle. Note, concentrate is not cooled.
High sugar syrups are more viscous (thicker) than diet syrups and consequently have more 
pressure drop within a system. This pressure drop results in less flow at the valve and therefore a 
slower fill time. Increasing the pump pressure will help overcome the pressure drop caused by 
high sugar syrups.
3. S
ETTING
 R
ATIOS
If the ratio varies and must be adjusted often, it is probably the result of a restricted syrup 
system. 
It is then time to clean and sanitize the syrup tubing and cooling coils and check for other 
problems such as syrup pumps, etc.
Note: Try raising the pressure on the pump before cleaning.
Set the water flow rate first, then adjust the syrup to the desired ratio. This gives the most 
accurate valve flow setting possible. Measure the ratio and adjust the syrup flow, if necessary. 
This will result in uniform flow, better carbonation retention, and improved drink quality.
When using a ratio cup always take these 
precautions to ensure accuracy:
• After installing the separator tube, open 
the valve to fill the syrup tube before 
starting the ratio test.
• Clean the cup thoroughly between tests 
so there is no carry–over from one test 
to the next.
• Fill the cup to approximately 3/4 full 
and use approximately the same quan-
tity for each test to ensure accurate set-
tings.
• Take another sample to verify the set-
tings.
Ratio
Cups
Syrup 
Separator
SY
ST
E
M
 D
E
TA
IL
S