TIC-LF494B 
 
 
2
Specifications 
 
Overall Specifications 
Measurement range in terms of flow velocity: 
0
 
–1.0
 
ft/s to 0
 
 
32.8
 
ft/s (0
 
 
0.3
 
m/s to 0
 
–10
 
m/s) 
0
 
 
0.3
 
ft/s to 0
 
 
1.0
 
ft/s (0
 
 
0.1
 
m/s to 0
 
 
0.3
 
m/s) 
range is available optionally. 
Accuracy: 
±
0.2 % of Rate
* 
* This pulse output error result is established under standard 
operating conditions at Toshiba's flow calibration facility, 
Fuchu Japan. (NIST Traceable).  
* Individual meter measurement error may vary up to ±0.5% of 
Rate at 1.64 ft/s (0.5m/s) or more and ±0.3% of rate ±0.039 
inch/s (1mm/s) at 1.64ft/s (0.5 m/s) or less. 
* Current output: plus ± 8µA (0.05% of span.) 
*Refer to individual calibration data for each individual meter's 
measurement error.
 
Fluid conductivity
: 5µS/cm minimum 
Fluid temperature:  
14 to 248 
°F (–10 to +120 °C) 
Ambient temperature: – 4 to 140 
°F (–20 to +60 °C) 
Structure: IP 67 and NEMA 4X Watertight 
Power consumption:  
17 W(27VA) or less 
19 W(29VA) or less (with PROFIBUS) 
Conformance to European Community Directives: 
EMC directive 89/336/EEC 
The low voltage 93/68/EEC 
3A standard (LF490 and LF494 detectors): 
Approved for 3A standard with FDA approved 
Teflon PFA lining. 
Approved hazardous location certifications: 
Model: LF494/LF610F and LF494/LF612F 
cFMus explosion proof: 
FM Class I, Division 2, Groups A,B,C, and D. 
FM Class II, Division 2, Groups E, F and G. 
FM Class III. 
Detector and converter combination: 
LF490/LF610  : Combined type for standard 
specification. 
LF490/LF612  : Separate type for standard 
specification. 
LF494/LF610F: Combined type with Ex approval 
of Class I, Division 2 (cFMus). 
LF494/LF612F: Separate type with Ex approval 
of Class I, Division 2 (cFMus). 
Mount-Anywhere
 Technology:   
Thanks to TOSHIBA’s unique patented magnetic 
field distribution technology, the meter is highly 
immune to upstream flow disturbances.  
A minimum of 1D (diameter) length of upstream 
straight pipe from the flange is required to 
maintain the performance specification. 
Note : The test results were obtained and demonstrated at 
TOSHIBA's flow calibration facility, Fuchu Japan. 
 
 
Model LF490 and LF494 Detectors 
Fluid pressure:   
–15 to 300 psi, or –1.0 to 20 bar (–0.1 to 2.0 MPa) 
Note:  This pressure is the allowable pressure for the 
detector. The actual pressure will be restricted by 
the type of connection method. For example, in 
the case of sanitary clamp type, the maximum 
pressure is 150 psi or 10 bar(1.0 MPa). 
Note  The test pressure before shipping from the 
factory is equal to twice the nominal pressure 
rating of the customer specified flange 
connection during 15 minutes. 
Connection method: 
Sanitary clamp type (ISO 2852) 
Note: 
  
Tri-clamp® can connect except meter size 4" (100 mm). 
  Tri-clamp® is a registered trademark for Tri-Clover Inc. 
Principal materials: 
Case — stainless steel 
Linings — Teflon PFA (FDA approved) 
Note:  Teflon liners are mechanically retained for 
full vacuum service. 
Electrodes — 316L stainless steel (std.) 
Note: Electrodes are electro-chemically polished 
after mechanically buffed.
 
Sanitary fittings — 304 stainless steel (std.) 
Seal gaskets — Silicon rubber (FDA approved) 
Note:   See Table 5 for optional materials and other 
related information. 
Measuring tube material — 304 stainless steel 
Coating: no coating (std.) 
Dimensions and weights: 
See Figures 3 to 6. 
Cable connection port: for separate type detectors. 
Cable glands — 
LF490: Provided as standard, R(PT) 1/2 
male screws. 
LF494: Not provided, 3/4–14NPT male 
screws are required. 
Applicable diameter —  
0.433 to 0.512 inch (11 to 13mm)