SECTION 1—IMPORTANT SAFEGUARDS
Part No 1150748
7
XPO
2
SECTION 1—IMPORTANT 
SAFEGUARDS
In order to ensure the safe installation, assembly and operation of the XPO
2
 Portable 
Concentrator these instructions MUST be followed.
 WARNING
SECTION 1 - IMPORTANT SAFEGUARDS contains important information for the safe 
operation and use of this product.
Operating Information
 DANGER
A spontaneous and violent ignition may occur if oil, grease, greasy substances, or 
petroleum based products come in contact with oxygen under pressure. These 
substances MUST be kept away from the XPO
2
 portable concentrator, tubing and 
connections, and all other oxygen equipment. DO NOT use any lubricants unless 
recommended by Invacare.
To Reduce The Risk Of Burns, Electrocution, Fire Or 
Injury To Persons
DO NOT come in contact with the concentrator while wet.
DO NOT place or store product where it can drop into water or other liquid.
DO NOT reach for product that has fallen into water. Unplug IMMEDIATELY.
Avoid creation of any spark near medical oxygen equipment. This includes sparks 
from static electricity created by any type of friction.
NEVER drop or insert any object or liquid into any opening.
For optimum performance, Invacare recommends that each concentrator be On, in auto-
pulse mode, and running for a minimum of 5 minutes. Shorter periods of operation may 
reduce maximum product life. Refer to Checking O
2
 Purity on page 14 for auto-pulse 
mode.
A product should NEVER be left unattended when plugged in. Make sure XPO
2
 is Off 
when not in use.
The XPO
2
 MUST be used in an upright position.
DO NOT connect the concentrator in parallel or series with other oxygen concentrators or 
oxygen therapy devices.