Shop-Vac XPO100 Manual De Usuario

Descargar
Página de 20
SECTION 2—COMPONENT REPLACEMENT
Part No 1150748
9
XPO
2
SECTION 2—COMPONENT 
REPLACEMENT
CAUTION
Ensure not to damage ribbon cable when removing or installing front cover.
DO NOT use Methyl Ethyl Ketone (MEK). This will damage the plastic.
Replacing the Gross Particle Filter or Filter Cover
NOTE: For this procedure, refer to 
FIGURE 2.1.
NOTE: The gross particle filter should be 
replaced as needed.
1. Lift filter cover slightly and pull down 
to remove tabs from the grooves.
2. Lift out the filter from the filter cover.
3. Use a vacuum cleaner or wash with a 
mild liquid dish detergent (such as 
Dawn
) and water. Rinse thoroughly. 
4. Thoroughly dry the filter and inspect 
for fraying, crumbling, tears and holes. 
5. Replace filter cover and/or filter.
6. Reinstall the filter cover by placing the 
tabs in the slots and engaging the 
magnet.
NOTE: Use only Invacare gross particle filter.
L
FIGURE 2.1   Replacing the Gross Particle 
Filter or Filter Cover
Accessing XPO
2
 WARNING
When accessing the XPO
2
 remember that battery power is still being applied to the 
internal (PCB) printed circuit board. Care MUST be taken to minimize contact with the 
PCB. DO NOT use any kind of tool (screwdriver, pliers, etc.) when working on the 
internal structure, otherwise, injury and/or damage may occur. Battery power remains on 
the PCB as long as the battery connection is made to the PCB. 
NOTE: For this procedure, refer to FIGURE 2.2 and FIGURE 2.3 on page 10.
Cover
Filter
Magnet
Cover
Tabs
Magnet
Grooves
Filter 
Cover