12.2  Manual control measurements 
FastFacts 12.2: Manual control measurements 
12.3  Sound level meter using REM probe 
FastFacts 12.3: Sound level meter using REM probe 
The manual control feature enables the RM500SL to be used as a sound level 
meter to perform C-weighted, A-weighted or 1/12
th
 octave analysis of sound as 
measured by the probe and reference microphones. A and C-weighted 
measurements are limited to frequencies above 100 Hz and the averaging time 
and update rate are 384 ms. 
1)  Refer to 
10: Real-Ear Measurement Setup
 to prepare the system for use and 
for proper positioning of the client and the probe tube. 
2)  Press <Tests>, then highlight & <PICK> [Manual control] in the REM column. 
3)  Highlight and <PICK> the Level window, then highlight and <PICK> 0 dB 
stimulus level from the drop-down list. 
4)  Highlight and <PICK> the Filter window, then highlight and <PICK> 1/12
th
 
octave, A-weighted, or C-weighted from the drop-down list. 
5)  For 1/12
th
 octave filter, highlight and <PICK> the Frequency window, then 
highlight and <PICK> the desired frequency from the drop-down list. 
6)  Highlight and <PICK> [Start test] to start the test. [Start test] changes to [Stop 
Test]. To stop the measurement, highlight and <PICK> [Stop test]. You may 
change Level and Frequency without stopping the test.  
1)  Refer to 
10: Real-Ear Measurement Setup
 to prepare the system for use and 
for proper positioning of the client and the probe tube. 
2)  Press <Tests>, then highlight and 
<PICK> [Manual control]. 
3)  Highlight and <PICK> the Level 
window, then highlight and <PICK> the 
desired stimulus level from the drop-
down list. 
4)  Highlight and <PICK> the Frequency 
window, then highlight and <PICK> the 
desired frequency from the drop-down 
list. (Distortion measurement data is 
available only from 200-2500 Hz.) 
5)  Highlight and <PICK> [Start test] to 
start the test. [Start test] changes to 
[Stop Test]. To stop the measurement, 
highlight and <PICK> [Stop test]. You 
may change Level and Frequency 
without stopping the test.