14  Occlusion Effect Test 
14.1  Occlussion effect measurement 
FastFacts 14.1: Occlusion effect measurement 
1)  Refer to 
10: Real-Ear Measurement Setup
 to prepare the system for use and 
for proper positioning of the client and the probe tube. 
2)  Insert the earmold or custom hearing instrument into the ear, being careful not 
to advance the probe tube further into the ear canal. A lubricant applied to the 
earmold or custom instrument shell in the vicinity of the probe tube will make 
insertion easier and prevent slit leaks around the tube. 
3)  Turn the hearing aid OFF. 
4)  Press <Tests>, then highlight and <PICK> [Occlusion]. 
5)  Highlight and <PICK> [Start test] to start the test. [Start test] changes to [Stop 
Test]. 
6)  While the client vocalizes “ee” in a normal voice, highlight and <PICK> [Stop 
test] to stop the test and freeze the screen. A red occlusion bar indicates an 
occlusion problem, a yellow bar indicates a possible problem and a green bar 
indicates no problem. 
7)  Increase venting and repeat the test until a green bar is obtained.