System Operation 
98
312779E
Purging Sampling Valves and Tubes
Follow this procedure after meter calibration.
1.
Go to Standby mode (remove Mix input).
2.
See F
IG
. 106, page 94. Close both fluid shutoff 
valves and sampling valves.
3.
Route the sampling tubes into a grounded waste 
container.
4.
On a single color system, attach a solvent supply 
line to Flow Meter A inlet.
5.
On the EasyKey, press Setup 
 key and access 
the Advanced Setup screens.
6.
Press the Right Arrow 
 key to select the Calibra-
tion screen. Press the Down Arrow 
 
key and 
select Purge from the menu. Press the Enter 
 
key
Dose A, solvent purge valve (B side), and color 
change solvent valves (if used) will open.
7.
To avoid splashing, slowly open the sampling 
valves and dispense solvent until the valves and 
tubes are clean.
NOTE: When performing a calibration purge, the 
solvent valve(s) close automatically after 2 minutes 
or when Abort is selected on the screen.
8.
Close sampling valves. 
NOTE: Select Abort on Calibration screen to cancel 
current calibration and close dose or purge valves.
9.
Fully open both fluid shutoff valves.
10. On a single color system, reconnect component A 
fluid supply line to flow meter A.
NOTE: After calibration it is necessary to clean out con-
taminated mix material. Do a manual purge and resume 
the recipe just tested, or do Recipe 0 then go on to the 
next recipe.
F
IG
. 107. Calibration Screen