Operation Mode – NOC
5.3.4
Other activity during Continuous mode measurement
During continuous measurement, you can use the 
EXIT
 button on the Well Performance Measurement 
screen to move between the continuous measurement screens and the View menu. From the View 
menu, you can perform any available actions.
At any point from the continuous measurement screens, you can press the Security button to access 
the Configuration and Maintenance menus. The system does not prevent configuration changes, 
calibrations, or other procedures. 
However, many actions (e.g., reconfiguring well data) will cause discontinuities in the data, and many 
procedures will interfere with data collection. If you need to change system configuration or perform 
a maintenance procedure:
1. Pause measurement.
2. If desired, save an archive record.
3. Make the required configuration changes or perform the required procedures.
4. Reset all process variables.
5. Resume measurement.
5.3.5
Viewing archive records
To view archive records, refer to Figure 5-5 and:
1. Select 
View Archives
 from the Well Performance Measurement menu. 
2. Specify the archive record you want to view.
3. Select 
Quick View
 or the specific process variable you want to view.